This October, SPL is spotlighting the specialty methods our technical services team runs every day. These methods take more than a standard instrument setup and a method number on the report. Each week we’ll walk through one method, the problem it was built to solve, and the people who make the results hold up.
We’re starting with GPA 2103, a method our team helped write. It answers a question every crude and condensate composition raises but rarely states on the report: what happened to the components too heavy for the column?
Key Points
- Every crude or condensate composition totals 100 percent, but the total can include heavy components that were never measured.
- Heavy hydrocarbons are too heavy to vaporize and elute on a GC column. Some methods normalize the observed components to 100 percent instead of accounting for the plus fraction.
- An undercounted heavy end shifts weight-to-liquid-volume conversions, which affects contract pricing and royalty calculations for up to a year of production.
- The same gap inflates the light end, which raises flash and flare estimates in emissions reports.
- GPA 2103 accounts for the plus fraction through additional analysis. SPL’s technical services team helped write the method and serves on the GPA Midstream committee that maintains it.
A compositional report looks the same whether it is complete or not. It carries no error flag and no footnote for the components a method could not measure. The number reads confidently either way, and that is where the cost begins.
Why Heavy Components Go Missing
A gas chromatography (GC) system relies on an entire sample being in the gas phase to be separated by the column and make it to the detector for quantification. Heavy hydrocarbon components, due to high molecular weight and thus higher temperatures required to vaporize, do not make it to the column and therefore cannot elute to the detector. Different standard methods treat this missing portion of the analysis in different ways, which can lead to reports that look complete but don’t show the whole picture. For example, GPA 2177 would have the component values that were observed normalized to equal 100%. However, GPA 2177 would not be applicable to such samples, as the plus fraction molecular weight is outside of the range that GPA 2177 allows.
The analysis is not wrong about what it reports, it stays silent about what it missed. A composition must total everything completely, so the weight that never eluted does not vanish. It lands somewhere else in the report, and can affect other important pieces of information beyond just the missing mass. If you do not know what is in the heavier components, you cannot accurately account for their densities and molecular weights and how those affect calculated overall properties and converting from your calibrated unit of measure to the other important concentration calculations, be it mole percent, weight percent, or liquid volume percent.
How GPA 2103 Handles the Heavy End
The Technical Services Team at SPL developed GPA 2103 at our laboratories around that limit instead of fighting it. Rather than trying to force heavy hydrocarbons to vaporize and move through the GC column, we instead account for them through further analyses. Key information about the heavy portion of the sample, often referred to as the plus fraction, is obtained by analysis of a weathered portion of the liquid sample. The weathered sample has its density measured, has its average molecular weight determined, and then is injected onto the GC again to account for any light hydrocarbons still present in the weathered portion
Cost One: A Stream Priced on the Wrong Composition
Commodity contracts price by composition. The relationship between a component and its price is not that simple. Heavier hydrocarbons are not always more valuable. What matters is that the composition feeding the price matches the stream you produced.
For example, your GC is calibrated in weight percent, but your contracts are based on liquid volume. The conversion from weight to liquid volume requires knowledge of that plus fraction density. If you apply a density based on constants for some components in the plus fraction, that will shift the composition. The price your contract applies then rests on a stream you did not produce.
Sampling frequency raises the stakes, as operators typically sample crude once or twice a year, not weekly. One analysis can stand in for a well’s entire year of production, and it feeds the royalty and revenue calculation until the next sample replaces it.
If that sounds like a contract your team reconciles every month, ask SPL to review a recent crude analysis.
Cost Two: Emissions You Report and Did Not Emit
The same gap works on the compliance side. If GPA 2177’s normalizing missing data across the components that were observed is used, the heavy components are under reported and the light components become over reported. Overstating the light components feeds directly into estimates of how much volume flashes to vapor in storage or to flares.
State agency emissions reports use those estimates, and some operators carry federal reporting obligations as well. An inflated estimate can mean a higher reported number than your facility produces.
A Systems Problem, Not a Single Sample Problem
Allocation and custody transfer across a gathering system depend on every point for reconciling: wellhead, gathering, storage, point of sale, etc. One weak composition skews volume and value balances at every point downstream of it. Accuracy does not mean any single number is right. It means the whole system balances.
A Five Question Check for your Current Lab
Use these with your current provider, whoever it is:
1. Which method is used on my crude or condensate samples?
2. How does that method treat components too heavy for the column?
3. How does it account for the weight of those components in the result?
4. Does my report state what the method could not measure?
5. Can the lab explain how the composition totals to 100 percent?
If a lab answers the second and fourth questions clearly, you have defensible data. If it hesitates, you have a gap to close.
What SPL brings to GPA 2103
SPL developed GPA 2103, shared that knowledge with GPA Midstream, and still serves on the committee that maintains it. The published method covers condensate samples with a hexanes plus molar mass over 93, per the GPA Midstream technical standards. Not only do we run this method, we helped write it.
Ready to see the heavy end in your own report? Send SPL a recent crude or condensate sample and we will run GPA 2103 so you can compare it with your current result. Request a GPA 2103 sample analysis.
Frequently Asked Questions
What is GPA 2103?
GPA 2103 is a published GPA Midstream gas chromatography method for analyzing condensate and crude mixtures. SPL developed it and shared it with GPA Midstream.
Why do heavy hydrocarbons go missing in a standard analysis?
A GC can only vaporize components to a certain weight. Components above its limit do not make it to the column and do not elute, so a method that does not compensate leaves them out of the result.
How does GPA 2103 account for the heavy end?
Rather than trying to force heavy hydrocarbons to vaporize and move through the GC column, we instead account for them through further analyses. The method backflushes to atmosphere the intermediate heavy components that are vaporized but struggle to elute from the column, then calculates the missing portion by difference for both heavy and intermediate components.
Does the report show what the lab could not measure?
No. Unmeasured components produce no error or flag. They are absent, so the composition looks complete.
How does an incomplete composition affect revenue?
Contracts price by composition. If the heavy end is undercounted, the composition behind your price no longer matches the stream you produced.
How often does an operator sample crude oil?
Typically once a year, far less often than a weekly or daily cadence. One analysis can therefore stand in for a well’s entire year of production.
How does it affect emissions reporting?
A composition must total completely, so an undercounted heavy end inflates the light end. That raises flash and flare estimates in storage and feeds state agency emissions reports.
Does SPL still work on GPA 2103?
Yes. SPL’s team serves on the GPA Midstream committee that maintains the method.
See the Heavy End in Your Own Report
Send SPL a recent crude or condensate sample. Our technical services team will run GPA 2103 so you can compare it side by side with your current result.